Solair BC Technikai adatok


Technical Details

  1100 LD 3100 5100 3350 Boulder Counter

Specifications

Size Range 0.1 - 1.0 μm
Optional ranges up to 5.0 µm
0.3 - 25.0μm 0.5 - 25.0μm 0.3 - 25.0 µm 5.0 - 100.0 um
Channel Sizes Standard 8 channels: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 0.7, 1.0μm
Optional 8 channels: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 1.0, 5.0 µm
Standard: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm
Standard: 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm
8-Channel: 0.3, 0.5, 0.7, 1.0, 3.0, 5.0,7.0,10.0 µm
Standard: 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 µm
Standard: 0.5, 1.0, 3.0, 5.0, 10.0, 25.0 µm
8-Channel: 0.5, 0.7, 1.0, 2.0, 3.0, 5.0, 7.0,10.0 µm
Standard: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm
Optional: 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm
Standard: 5.0, 10.0, 25.0, 40.0, 50.0,100.0µm 
Flow Rate 1.0 CFM (28.3 LPM) 100 LPM 28.3 LPM
Counting Efficiency 50% @ 0.1 µm; 100% for particles > 0.15 µm per ISO 21501-4 50% @ 0.3 µm; 100% @ sizes
> 0.45 µm
50% @ 0.5 um; 100% @ sizes
> 0.75 µm
50% @ 0.3 µm; 100% @ sizes > 0.45 µm 50% @ 5.0 um
Concentration Limits 500,000 Particles / ft3 (17,150,000 / m3) @ 5% Coincidence Loss 1,000,000 particles/ft3
(35,300,000/m3) @ 10% coincidence loss
Suitable for ISO classes 1-8 cleanrooms
400,000 particles/ft3 (14,120,000/m3)
@ 10% coincidence loss
Suitable for ISO classes 1-8 cleanrooms
500,000 particles/ft3 @ 5% coincidence loss 
Light Source Long Life Laser Diode Extreme Life Laser Diode Technology with >20-year Mean Time To Failure (MTTF)
Dimensions 14.65"Lx8.0"Wx10.4"H (37x20x26cm) 9.5"Lx8.0"Wx10.2"H (24.13x20.32x25.9cm) 14.65"Lx8.0"Wx10.4"H (37x20x26cm) 9.5"Lx8.0"Wx10.2"H (24.13x20.32x25.9cm)
Weight With battery: 23.55 lbs (10.68 kg)
Without battery: 20.45 lbs (9.28 kg)
With battery: 13.40 lbs (6.08 kg)
Without battery: 12.00 lbs (5.44 kg)
With battery: 26.15 lbs (11.86 kg)
Without battery: 23.05 lbs (10.46 kg)
With battery: 13.40 lbs (6.08 kg)
Without battery: 12.00 lbs (5.44 kg)
Zero Count Level < 1 count / 5 minutes (per JIS B 9921)
Calibration Meets ISO 21501-4 Calibration using NIST Traceable PSL Spheres, DMA and Condensation Particle Counter
Count Modes Automatic, Manual, Beep, Concentration, Cumulative/Differential, Real Time (with LMSNet or LMS Express RT software)
Data Storage 3,000 Sample Records, Rotating Buffer (particle and environmental data with location and time)
Communication Modes USB, Ethernet, RS-485 and Secure Data Transfer (LSDT) to and Storage on USB Flash Drive
Supporting Software LMS Express and Express RT; LMS Pharma
Environmental Sensors Optional (up to 4) temperature/relative humidity, air velocity, differential pressure or others
Touch Screen Display 5.7-inch (14.47 cm) Color TFT Touch Screen Display
Reports Pass/Fail Report Wizard for ISO-14644-1, EU GMP Annex 1 and FS-209 Report Formats
Alarms Integrated alarm buzzer and external alarm connector. Alarms on counts, low battery, sensor failure
Sample Output Internally Filtered to 100% @ 0.1μm, Optional HEPA Filtration for Cooling System Exhaust Internally filtered to HEPA standards (>99.97% @ 0.3μm)
Vacuum Source Internal closed-loop controlled pump. Allows the use of 3/8" ID sample tubing of up to 12 meters (40 feet) @ 50 & 100LPM
Enclosure Stainless Steel
Power 100-240 VAC, 50-60 Hz, External AC/DC Adapter
Battery Optional: Li-Ion, Removable & Rechargeable (typical recharge time 7.5 hours)
Languages English (U.S.), French, German, Italian, Spanish, Chinese, Russian, Japanese, Korean

Environmental Conditions

Operating 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing
Storage 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing

Accessories

Included Operating Manual on CD; Isokinetic Sample Probe with Tripod and Tubing; Purge Filter; Power Cable and Spare Fuse; LMS XChange and Express Software; Printer; Printer Paper
Optional Printed Operating Manual; Network Adapter; Wireless Network Adapter; Internal Removeable, Rechargeable Battery; Carrying Case with Wheels; Cleanroom Cart; Temperature/Relative Humidity Probe; Air Velocity Probe; Differential Pressure Probe; Filter Scanning Probe; High Pressure Difuser; 32-Port Universal Manifold; LMS Express RT Software; LMSNet Software; Validation Documentation Wireless Network Adapter; Removable Rechargeable Li-Ion Battery; Carrying Case with Wheels; Temperature/Relative Humidity Probe; Differential Pressure Probe; Intelligent Filter Scanning Probe; Air Velocity Probe; Spare Removable
Rechargeable Li-Ion Battery; External Battery Charger; LMS Express, RT Software; LMS Pharma Edition Software; LMS Standard Software; Validation Documentation; Printed Operating Manual; Cleanroom Cart

A SOLAIR Boulder Counter magában foglalja az Extreme Life Laser Diode Technology érzékelőnket, hogy az iparág leghosszabb, 20+ éves MTTF lézerdiódájának élettartamát állítsa elő (folyamatos, 24 órás működés alapján).

Az új, könnyebb, hosszabb élettartamú akkumulátor és külső AC adapter használatára tervezett SOLAIR Boulder Counter még hordozhatóbb. A SOLAIR Boulder Counter az iparág egyik legjobb felhasználói felületével (UI) rendelkezik a műszer konfigurálásához és működtetéséhez, intuitív ICON-vezérelt képernyőkkel és menükkel, amelyek gyorsan dokumentálhatók az SOP-hoz, könnyen megtanulhatók és egyszerűen használhatóak. A SOLAIR Boulder Counter akár 3000 rekordot is képes tárolni a részecskeszámlálási adatokról akár 6 részecskecsatornáról, a konfigurálható receptadatbázis pedig akár 50 receptet is tárolhat mintavételhez és a riportokhoz. Minden adat gyorsan és megbízhatóan átvihető számítógépre, létesítményfelügyeleti rendszerre vagy épületautomatizálási rendszerre. Biztonságos adatátvitel (21 CFR Part 11 szerint) USB flash meghajtóra a Lighthouse Secure Data Transfer funkció és az LMS Express vagy Express RT szoftver használatával valósítható meg. Ezekkel a műszerekkel való rugalmas kommunikációt az integrált Ethernet, RS-485, USB és USB Flash Drive interfészek teszik lehetővé. A SOLAIR Boulder Counter olyan szabványok szerint lett tervezve, gyártva és kalibrálva, amelyek biztosítják a részecskeszámlálás pontosságát és megismételhetőségét.

 

Főbb jellemzők:

  • 0,5 µm -100,0µm
  • 1,0 CFM (28,3 LPM) áramlási sebesség
  • Simultán akár 6 szemcseméret ellenőrzése
  • 21 CFR Part 11 kompatibilis - Biztonságos adatátvitel USB flash meghajtón keresztül
  • 20 év MTTF lézer élettartam
  • Thermal printer
  • Rozsdamentes acél szerkezet
  • Jelentésvarázsló tisztatér-megfelelés/hiba jelentésekhez: FS-209E, ISO 14644-1 (2015-ös verzió) és EU GMP 1. melléklet
  • Több kommunikációs interfész: USB, RS-485, Ethernet és USB flash meghajtó
  • 5,7 hüvelykes színes TFT érintőképernyő
  • 3000 rekord adattároló memória
  • 50 rekordos konfigurálható eljárás adatbázis
  • Open Architecture Connectivity with OPC Server
  • Cleanroom Compliance Reporting Wizard / Error Reports: FS-209E, ISO-14644-1 or EU GMP Annex 1
  • Több nyelv támogatása: angol, francia, német, olasz, spanyol, kínai, orosz, japán, koreai

 

 

Termék fileok


    Brochure – Solair 1100LD, 3100, 3350, 5100, Boulder Counter  Letöltés
    Data Sheet – Solair 3100, 3350, 5100, Boulder Counter  Letöltés
    Operating Manual – Solair 3100, 5100, Boulder Counter Gen E  Letöltés

 


Kapcsolódó Termékek

 

Solair 1100LD

  • 0,1 – 5 μm részecske méretű csatornák
  • 1 CFM áramlási sebesség

Solair 3350

  • 0,3 – 25 μm részecske méretű csatornák
  • 1 CFM áramlási sebesség

Solair 3100

  • 0,3 – 25 μm részecske méretű csatornák
  • 1 CFM áramlási sebesség

Solair 5100

  • 0,5 – 25 μm részecske méretű csatornák
  • 1 CFM áramlási sebesség
Olvass tovább >  Olvass tovább >  Olvass tovább >  Olvass tovább >